The MCS300P HW is a hot/extractive analyzer system for measurement of gases. It measures IR- or VIS-active components with variable measuring ranges for monitoring and controlling of combustion and flue gas conditioning. The heated system components prevent a temperature drop below the water and acid dewpoint. This results in lower maintenance and higher reliablity and long-time stablity of the system, also under rugged and demanding conditions. Automatic adjustment functions, easy operation concept and modern communication protocols results in a quick and easy integration into a local control system network.
At a glance
Simultaneous measurement of up to 6 components plus O2 |
Sample flow control and sample gas pressure measurement |
Temperature of heated system components up to 220 °C |
Automatic sample point switching for up to 8 sample points (option) |
Automatic adjustment of zero and span point |
Integrated adjustment unit without span gas (option) |
Extended operation via PC and software |
Flexible I/O modules |
Your benefits
Economic, automatic adjustment without expensive test gases |
Easy remote control by integration in existing networks |
Automatic control of the complete measuring system and probe |
Low maintenance and reliable due to hot measurement |
In combination with probe SCP3000 also for measurements at high dust loads and high temperatures |
Short response time due to high sample gas flow |
Fields of application
Monitoring of flue gas conditioning in waste or sewage sludge incinerators |
Monitoring of the scrubber in flue gas conditioning |
Measurement directly at the kiln inlet in cement plants |
Technical data | |||
Analyzer: | MCS300P | ||
Measurement principles: | Gas filter correlation, Interference filter correlation, Zirconium dioxide sensor | ||
Measured values: | CO, CO2, H2O, HCl, N2O, NH3, NO, NO2, O2, SO2 | ||
Max. number of measurands: | 6 | ||
Measuring ranges: | CO: 0 ... 800 ppm / 0 ... 5 Vol.-% CO2: 0 ... 20 Vol.-% / 0 ... 100 Vol.-% H2O: 0 ... 1 Vol.-% / 0 ... 60 Vol.-% HCl: 0 ... 700 ppm / 0 ... 12,000 ppm N2O: 0 ... 500 ppm / 0 ... 7,500 ppm NH3: 0 ... 550 ppm / 0 ... 5,000 ppm NO: 0 ... 750 ppm / 0 ... 15,000 ppm NO2: 0 ... 500 ppm / 0 ... 4,000 ppm O2: 0 ... 1 Vol.-% / 0 ... 21 Vol.-% SO2: 0 ... 200 ppm / 0 ... 10 Vol.-% |
||
Comment: | 2 measuring ranges per component, Other measuring ranges and components on request | ||
Gas flow rate: | 150 ... 800 l/h | ||
Температура процесса: | Sample gas inlet: ≤+220 °C Process: ≤+1,300 °C depending on sampling probe |
||
Давление процесса: | 800 ... 1,200 hPa | ||
Ambient temperature: | +5 ... +35 °C temperature change ±10 °C/h maximum With cooling device: +5 ... +45 °C temperature change ±10 °C/h maximum; other temperatures on request |
||
Ambient humidity: | ≤80 % non-condensing |
||
Electrical safety: | CE | ||
Enclosure rating: | IP 43 versions with NOx-converter (option) |
||
Analog outputs: | 2 outputs: 0/4 ... 22 mA, 500 Ω electrically isolated; additional outputs if using I/O modules (option) |
||
Analog inputs: | 2 inputs: 0/4 ... 22 mA, 100 Ω electrically isolated; additional inputs if using I/O modules (option) |
||
Digital outputs: | 2 power relays electrically isolated; additional outputs if using I/O modules (option) 3 potential-free contacts additional outputs if using I/O modules |
||
Digital inputs: | 2 potential-free contacts additional inputs if using I/O modules |
||
Interfaces: | Ethernet | ||
Bus protocol: | Modbus TCP OPC |
||
Operation: | Application specific programs for system control, Via operating unit in the analyzer, Via software SOPAS ET, Two operating levels, one password-protected | ||
Device version: | Steel sheet cabinet | ||
System components: | Analyzer cabinet Sampling probe Sample gas line, heated Sample point switching (max. 8 sample points) |
||
Test functions: | Automatic check cycle for zero and span point | ||
Corrective functions: | Internal adjustment unit (option) Cross-sensitivity compensation of up to 6 interferents |