Analyzer system with sample gas cooler, MCS100E CD

Главная » Продукция » SICK » Анализаторы » Analyzer solutions » Analyzer system with sample gas cooler, MCS100E CD
 
 

The MCS100E CD is an analyzer system for extractive measurement of up to 8 IR-active gas compounds. It can be supplemented with an oxygen sensor. The gas is dried by an integrated gas cooler allowing measurement of very small measuring ranges, especially for SO2, NO and NO2.

At a glance

  • Extractive measurement of up to 8 IR-active gas compounds
  • Optional oxygen sensor available
  • Integrated gas cooler
  • Test gas supply at the gas sampling probe or at the analyzer
  • Back-purging of gas sampling probe for cleaning of filters
  • Fast sample gas exchange for minimizing adsorption and desorption effects
  • Automated sample point switching

    Your benefits

  • Especially low measurement ranges for SO2, NO, NO2
  • Selective measurement of NO and NO2 - no converter required
  • Measurement of several gas components with one analyzer

    Fields of application

  • Flue gas monitoring in power stations
  • Monitoring of combustion plants
  • Emission monitoring in industrial plants
    Technical data
    Analyzer: MCS100E
    Measurement principles: Gas filter correlation, Interference filter correlation, Zirconium dioxide sensor
    Measured values: CH4, CO, CO2, H2O, N2O, NO, NO2, O2, SO2
    Max. number of measurands: 8
    Measuring ranges: CH4: 0 ... 140 ppm / 0 ... 1,400 ppm
    CO: 0 ... 40 ppm / 0 ... 2,000 ppm
    CO2: 0 ... 25 Vol.-% / 0 ... 100 Vol.-%
    H2O: 0 ... 1 Vol.-% / 0 ... 5 Vol.-%
    N2O: 0 ... 50 ppm / 0 ... 1,000 ppm
    NO: 0 ... 40 ppm / 0 ... 1,900 ppm
    NO2: 0 ... 40 ppm / 0 ... 500 ppm
    O2: 0 ... 1 Vol.-% / 0 ... 21 Vol.-%
    SO2: 0 ... 4 ppm / 0 ... 5,000 ppm
    Comment: Other measuring ranges and components on request, Measuring ranges depend on application and device version
    Gas flow rate: 200 ... 1,000 l/h
    depending on application
    Температура процесса: Inlet analyzer system: 0 ... +220 °C
    Process: 0 ... +1,300 °C
    Давление процесса: 900 ... 1,100 hPa
    atmospheric
    Ambient temperature: +5 ... +35 °C
    With cooling device: +5 ... +50 °C
    Ambient pressure: 900 ... 1,100 hPa
    Conformities: GOST
    MCERTS
    Electrical safety: CE
    Enclosure rating: IP 43
    Analog outputs: 4 outputs:
    0/4 ... 20 mA, 500 Ω
    max. 64 outputs with opto box via optical fibre possible
    Analog inputs: 4 inputs:
    0/4 ... 20 mA
    max. 64 inputs with opto box via optical fibre possible
    Digital outputs: 12 outputs:
    40 V, 1 A
    max. 64 outputs with opto box via optical fibre possible
    Digital inputs: 4 inputs:
    50 V AC, 4 A 24 V DC, 4 A
    max. 64 inputs with opto box via optical fibre possible
    Interfaces: Modem, analog
    additional interfaces on request
    RS-232
    RS-485
    Bus protocol: Modbus
    additional buses on request
    Operation: Sequence programs can be programmed as required, Via integrated operating unit, Two operating levels, one password-protected
    System components: MCS100E analyzer
    Analyzer cabinet
    Sampling probe
    Sample gas cooler
    Sample gas line
    Sample gas pump
    Test functions: Automatic check cycle for zero and span point
    Internal calibration filter for QAL3 drift check without test gas (option)
     
     

     
     
     