+7 (812) 432-38-38
+7 (812) 432-36-36
Пн-Пт 9:00 — 18:00
 
Заказать оборудование
 

Ship emission measuring devices, MARSIC

Главная » Продукция » SICK » Анализаторы » Analyzer solutions » Ship emission measuring devices, MARSIC
 
 

SICK’s MARSIC ship emission measuring device is an innovative solution that can be used to reliably measure emissions on ships in accordance with MARPOL Annex VI and MEPC.184(59). MARSIC has been given DNV GL type approval for this purpose and can be used to measure SOx and CO2 after use of scrubber, as well as NOx before and after use of SCR (Selective Catalytic Reduction) systems. As a leading manufacturer of emission measurement technology, SICK’s MARSIC offers tried-and-tested measurement technology within an impressive modular housing concept. Designed to be used on ships, it boasts extremely straightforward maintenance, and modules that are easy to replace.

At a glance

  • Rugged design and high level of measurement accuracy
  • Up to four measurement points with one analyzer
  • DNV GL type approval in accordance with MARPOL Annex VI, NOx Technical Code 2008 and MEPC.184(59)
  • Up to five measuring components at the same time: SO2, CO2, NO, NO2 and O2

    Your benefits

  • Low costs due to minimal maintenance requirements
  • Several measurement points with just one analyzer
  • High levels of availability and reliability
  • Assemblies that are easy to replace during servicing
  • Service support provided worldwide
  • Quick access to expert support with remote maintenance
  • Equipped for existing and future requirements thanks to its rugged design and high level of measurement accuracy

    Fields of application

  • Reliable emission monitoring on ships
  • Accurate DeNOx and DeSOx measurements for ship engines
  • Cost-efficient process monitoring downstream of the scrubber and catalytic converters
    Technical data
    Measurement principles: Electrochemical cell, NDIR spectroscopy, NDUV spectroscopy
    Measured values: CO2, NO, NO2, O2, SO2
    Max. number of measurands: 5
    Measuring ranges: CO2: 0 ... 25 Vol.-%
    NO: 0 ... 300 ppm / 0 ... 1,500 ppb
    NO2: 0 ... 200 ppm / 0 ... 500 ppm
    O2: 0 ... 21 Vol.-%
    SO2: 0 ... 100 ppm / 0 ... 500 ppm
    Температура процесса: +10 ... +550 °C
    Давление процесса: -90 ... 200 hPa
    Dust load: ≤200 mg/m3
    Ambient temperature: +5 ... +45 °C
    Ambient pressure: 900 ... 1,100 hPa
    Ambient humidity: ≤90 %
    non-condensing
    Conformities: DNV GL Rules for Type Approvals (2012)
    Guidelines for exhaust gas cleaning systems - MEPC.184(59)
    MARPOL Annex VI and NTC 2008 - MEPC.177(58)
    Electrical safety: CE, DNV GL Rules
    Enclosure rating: IP 54
    Analog outputs: 8 outputs:
    0 ... 24 mA
    electrically isolated
    Analog inputs: 2 inputs:
    0 ... 20 mA
    Digital outputs: 8 outputs
    changeover switch, 1-pin, 3 connections
    Digital inputs: 8 inputs:
    42 V
    Interfaces: Ethernet TCP/IP
    Bus protocol: Modbus TCP
    OPC
    Operation: Via LC display or SOPAS ET software, multiple operating levels, password-protected
    Device version: Steel sheet enclosure
    Corrective functions: Automatic zero point adjustment with N2, weekly
    Mounting: Wall-mounting
     
     

     
     
     