The MCS100E PD is an analyzer system for extractive measurement of up to 8 IR-active gas compounds. It can be supplemented with an oxygen sensor. The gas is dried by a permeation dryer, whereby HCl remains in the gas. Very small measuring ranges are possible, especially for SO2, NO, NO2 and HCl.
At a glance
Extractive measurement of up to 8 IR-active gas compounds |
Additional oxygen sensor as an option |
Integrated permeation dryer |
Test gas supply at the gas sampling probe or at the analyzer |
Back-flushing of gas sampling probe for cleaning of filters |
Fast sample gas exchange for minimizing adsorption and desorption effects |
Automated sample point switching |
Your benefits
Especially low measurement ranges for SO2, NO, NO2 and HCl |
Selective measurement of NO and NO2 - no converter required |
Measurement of several gas components with one analyzer |
Long maintenance intervals (typically 6 months) due to self monitoring of the analyzer |
Fields of application
Flue gas monitoring in waste incineration plants |
Clean gas monitoring in industrial plants |
Emission monitoring in refineries |
Technical data | |||
Analyzer: | MCS100E | ||
Measurement principles: | Gas filter correlation, Interference filter correlation, Zirconium dioxide sensor | ||
Measured values: | CH4, CO, CO2, H2O, HCl, N2O, NO, NO2, O2, SO2 | ||
Performance-tested measurands: | CO, CO2, HCl, NO, NO2, O2, SO2 | ||
Max. number of measurands: | 8 | ||
Measuring ranges: | CH4: 0 ... 140 ppm / 0 ... 1,400 ppm CO: 0 ... 40 ppm / 0 ... 2,000 ppm CO2: 0 ... 25 Vol.-% / 0 ... 100 Vol.-% H2O: 0 ... 1 Vol.-% / 0 ... 5 Vol.-% HCl: 0 ... 6 ppm / 0 ... 1,900 ppm N2O: 0 ... 50 ppm / 0 ... 1,000 ppm NO: 0 ... 40 ppm / 0 ... 1,900 ppm NO2: 0 ... 40 ppm / 0 ... 500 ppm O2: 0 ... 1 Vol.-% / 0 ... 21 Vol.-% SO2: 0 ... 4 ppm / 0 ... 5,000 ppm |
||
Comment: | Other measuring ranges and components on request, Measuring ranges depend on application and device version | ||
Gas flow rate: | 200 ... 1,000 l/h depending on application |
||
Температура процесса: | Inlet analyzer system: 0 ... +220 °C Process: 0 ... +1,300 °C |
||
Давление процесса: | 900 ... 1,100 hPa atmospheric |
||
Ambient temperature: | +5 ... +35 °C With cooling device: +5 ... +50 °C |
||
Ambient pressure: | 900 ... 1,200 hPa | ||
Conformities: | 2000/76/EC 2001/80/EC EN 14181 EN 15267 GOST MCERTS U.S. EPA compliant Approved for plants requiring approval |
||
Electrical safety: | CE | ||
Enclosure rating: | IP 43 | ||
Analog outputs: | 4 outputs: 0/4 ... 20 mA, 500 Ω max. 64 outputs with opto box via optical fibre possible |
||
Analog inputs: | 4 inputs: 0/4 ... 20 mA max. 64 inputs with opto box via optical fibre possible |
||
Digital outputs: | 12 outputs: 40 V, 1 A max. 64 outputs with opto box via optical fibre possible |
||
Digital inputs: | 4 inputs: 50 V AC, 4 A 24 V DC, 4 A max. 64 inputs with opto box via optical fibre possible |
||
Interfaces: | Modem, analog additional interfaces on request RS-232 RS-485 |
||
Bus protocol: | Modbus additional buses on request |
||
Operation: | Sequence programs can be programmed as required, Via integrated operating unit, Two operating levels, one password-protected | ||
System components: | MCS100E analyzer Analyzer cabinet Sampling probe Sample gas line Sample gas pump Permeation dryer |
||
Test functions: | Automatic check cycle for zero and span point Internal calibration filter for QAL3 drift check without test gas (option) |