The MCS100E HW is an analyzer system for extractive measurement of up to 8 IR-active gas compounds. It can be supplemented with oxygen and total hydrocarbon analyzers. From probe sampling to cell, all components that are in contact with the sample gas are heated above dew point and thus protected from corrosion. This hot measurement technique allows the measurement of water soluble components HCl and NH3.
At a glance
Extractive measurement of up to 8 IR-active gas compounds |
Additional oxygen and total hydrocarbon analyzers as an option |
Gas paths completely heated |
Test gas supply at the gas sampling probe or at the analyzer |
Back-purging of gas sampling probe for cleaning of filters |
Fast sample gas exchange for minimizing adsorption and desorption effects |
Automated sample point switching |
Your benefits
Measurement of several gas components with one analyzer |
Heated gas paths enables measurement of difficult gases like HCl and NH3 |
Long maintenance intervals (typically 6 months) due to self monitoring of the analyzer |
Selective measurement of NO and NO2 - no converter required |
QAL3 drift test according to EN 14181 with internal calibration filter wheel - no test gas required |
Fields of application
Emission monitoring of waste incineration plants |
Monitoring of combustion plants with secondary fuels |
Flue gas monitoring in cement plants |
Emission monitoring of refineries |
Emission monitoring of heavy oil fired diesel engines |
Technical data | |||
Analyzer: | MCS100E | ||
Measurement principles: | Gas filter correlation, Interference filter correlation, Zirconium dioxide sensor | ||
Measured values: | CH4, CO, CO2, H2O, HCl, N2O, NH3, NO, NO2, O2, SO2 | ||
Performance-tested measurands: | CO, CO2, H2O, HCl, NH3, NO, O2, SO2 | ||
Max. number of measurands: | 8 | ||
Measuring ranges: | CH4: 0 ... 70 ppm / 0 ... 1,400 ppm CO: 0 ... 60 ppm / 0 ... 2,000 ppm CO2: 0 ... 25 Vol.-% / 0 ... 100 Vol.-% H2O: 0 ... 1 Vol.-% / 0 ... 40 Vol.-% HCl: 0 ... 10 ppm / 0 ... 1,900 ppm N2O: 0 ... 50 ppm / 0 ... 1,000 ppm NH3: 0 ... 30 ppm / 0 ... 660 ppm NO: 0 ... 150 ppm / 0 ... 1,900 mg/m3 NO2: 0 ... 50 ppm / 0 ... 500 ppm O2: 0 ... 1 Vol.-% / 0 ... 21 Vol.-% SO2: 0 ... 25 ppm / 0 ... 5,000 ppm |
||
Comment: | Other measuring ranges and components on request, Measuring ranges depend on application and device version | ||
Gas flow rate: | 200 ... 1,000 l/h depending on application |
||
Температура процесса: | Inlet analyzer system: 0 ... +220 °C Process: 0 ... +1,300 °C |
||
Давление процесса: | 900 ... 1,100 hPa atmospheric |
||
Ambient temperature: | +5 ... +35 °C With cooling device: +5 ... +50 °C |
||
Ambient pressure: | 900 ... 1,100 hPa atmospheric |
||
Conformities: | 2000/76/EC 2001/80/EC EN 14181 EN 15267 GOST MCERTS U.S. EPA compliant Approved for plants requiring approval |
||
Electrical safety: | CE | ||
Enclosure rating: | IP 43 | ||
Analog outputs: | 4 outputs: 0/4 ... 20 mA, 500 Ω max. 64 outputs with opto box via optical fibre possible |
||
Analog inputs: | 4 inputs: 0/4 ... 20 mA max. 64 inputs with opto box via optical fibre possible |
||
Digital outputs: | 12 outputs: 40 V, 1 A max. 64 outputs with opto box via optical fibre possible |
||
Digital inputs: | 4 inputs: 50 V AC, 4 A 24 V DC, 4 A max. 64 inputs with opto box via optical fibre possible |
||
Interfaces: | Modem, analog additional interfaces on request RS-232 RS-485 |
||
Bus protocol: | Modbus additional buses on request |
||
Operation: | Sequence programs can be programmed as required, Via integrated operating unit, Two operating levels, one password-protected | ||
System components: | MCS100E analyzer Analyzer cabinet Sampling probe Sample gas line Sample gas pump |
||
Test functions: | Automatic check cycle for zero and span point Internal calibration filter for QAL3 drift check without test gas (option) |